BALANCES 3Y SERIES
ultra-microbalances UYA 3Y
analytical balances XA 3Y
analytical balances XA 3Y.A
analytical balances AS 3Y
precision balances PS 3Y
precision balances APP 3Y
BALANCES AND SCALES
RADWAG 26 600 Radom, Bracka 28, POLAND
Phone: +48 (0-48) 38 48 800, fax. +48 (0-48) 385 00 10
export@radwag.com
www.radwag.com
Page view 0
1 2 3 4 5 6 ... 190 191

Summary of Contents

Page 1 - BALANCES 3Y SERIES

User manual no.: LMI-59-06/12/12/ENG BALANCES 3Y SERIES − ultra-microbalances UYA 3Y − analytical balances XA 3Y − analytical bala

Page 2

- 10 - 2. UNPACKING AND INSTALLATION 2.1. Ultra-microbalaces UYA 3Y After ultra-microbalance installing at operating place, following should be s

Page 3

- 100 - Procedure: • Enter submenu < Databases>, press < Pipettes> field • Press < Add > soft key, should a new pipette be add

Page 4

- 101 - • On setting correct values return to the main window. 24.4. Printouts Option Printouts enables setting the content of a standard printout

Page 5

- 102 - 24.5. Working mode activating procedure Carrying out pipettes calibration process requires installing an adapter for pipettes calibration. T

Page 6

- 103 - If option <Acquire ambient conditions from THB module> is enabled (set to <YES>, the software automatically acquires data from th

Page 7

- 104 - An example of a report from pipette calibration process: ----------- Pipette calibration ----------- User

Page 8 - 1. BASIC INFORMATION

- 105 - 25. DIFFERENTIAL WEIGHING Working mode < Differential weighing> enables analyzing changes of mass in a single sample or multiple s

Page 9

- 106 - On selecting the differential weighing mode the display contains the following quick access keys in the bottom bar: 1. Setup – access to

Page 10

- 107 - by selecting other quick access keys to the bottom bar of the display. Such process requires appropriate operator’s access level. Below list

Page 11

- 108 - 25.3. Introducing a series to the Database of Series Database of series comprises series and samples which are components of a series. Wh

Page 12

- 109 - SELECTING a series: Press quick access key <SERIES>. A window is displayed with database of series. On entering the database, press f

Page 13 - CAUTION:

- 11 - 2.2. Balances XA 3Y; XA 3Y.A XA 3Y analytic balances A – balance with d=0,01mg unit B – balance with d=0,1mg unit After balance installing

Page 14

- 110 - Weighing A On selecting an option, the balance moves to the first sample on the list, for which weighing <A> is still to be carried ou

Page 15 - 3. BALANCE DESIGN

- 111 - Follow above procedure and weigh the following samples. The process can be aborted by pressing < > key. Weighing T On selecting the

Page 16

- 112 - Caution: If option <Ambient conditions> in the working mode is disabled – set to <No>, the software skips this part of the proce

Page 17

- 113 - The opened window contains the following data: In case the following procedure is completed for a sample: • weighing A only, then tare va

Page 18

- 114 - Remember, that the software requests to carry out the <FIRST> cycle out of <3> from all samples in a series, for which this measu

Page 19 - 4 – IN / OUT, RS232 (COM2)

- 115 - On previewing a specific weighing, a window is opened with data and values on the weighing process. Numbers displayed next to weighing data &

Page 20 - 4. STARTUP

- 116 - which opens a list of samples for which the selected procedure can be carried out. Select a sample to be weighed from displayed list. This op

Page 21 - 6. SOFTWARE STRUCTURE

- 117 - Content of a printout of a series: − Dashes − Client − Samples − Empty line − Dashes − Signature − Non-standard printout Content of

Page 22 - 7. SOFTWARE WEIGHING WINDOW

- 118 - 26. STATISTICAL QUALITY CONTROL – SQC Working mode < Statistical Quality Control > is intended to carrying out different types of pro

Page 23 - 8. LOGGING

- 119 - 26.1. Working mode activating procedure Procedure: • while in the main window, press soft key located in the upper bar of the display, w

Page 24

- 12 - 2.3. Balances AS 3Y series Carefully remove the balance from its packaging, remove the plastic and foil transport protective elements. Gentl

Page 25 - 9. MOVING IN USER MENU

- 120 - Caution: Data on a product, listed in the below table, have to be filled in before starting a control: Mase Nominal mass of a product Ta

Page 26

- 121 - 26.2. Additional settings of the SQC mode The additional settings enable adjusting the working mode to user’s needs and requirements. Acc

Page 27 - 10. WEIGHING

- 122 - Messages displayed during a control process: Batch quantity 1 / 10 - A command on process course and the quantity of all measurements in a t

Page 28

- 123 - Net 50.053 g ---------Measurement 6 --------------- Net 50.030 g

Page 29

- 124 - 27. CONTROL OF PREPACKAGED GOODS (mode not available in balance’s standard version) Working mode < Control of Prepackaged Goods > ena

Page 30

- 125 - Control start: • Starting a control requires logging in a user with corresponding access level. Caution: 3. Starting a control requires t

Page 31 - II > pictogram on its

- 126 - Caution: Editing product records in the database is described in point 27.2 of this user manual. • After returning to the main window of th

Page 32

- 127 - − Activating average tare determining mode Before starting control type “non-destructive with average tare” the balance user can determine

Page 33 - !> symbol

- 128 - Caution: In order to carry out a control using the “non-destructive with average tare” mode, the standard deviation “S” of packaging mass, de

Page 34 - • SINGLE

- 129 - During control process the software on an on-going basis analyses the measurement results and displays them in the corresponding fields to in

Page 35

- 13 - 2.5. Balances APP 3Y series Gently take the balance out of its packaging, remove plastic, carton, foil and transport protections, and gently

Page 36 - • NON-STANDARD PRINTOUT

- 130 - Pressing the causes printing a report on a printer connected to the balance. Pressing the key causes returning to the main window of the

Page 37 - 11. ADJUSTMENT

- 131 - Messages and commands displayed in “Non-destructive empty-full” control: Place empty 1 / 5 - A command on process course and the value of th

Page 38

- 132 - On control completion, the software generates a process summary and the completed control is automatically saved in the balance’s database.

Page 39

- 133 - 26.7. A report from product control An example of a report: ----- CPG Report: U/22/08/12/14/22 ----- --------

Page 40 - Hint:

- 134 - 27. DATABASES Software of balances 3Y series features the following databases < >: PRODUCTS WEIGHING RECORDS CLIENTS FORMULAT

Page 41 - 12. USERS

- 135 - 27.3. Processes carried out on databases Databases < > are accessed by pressing key or corresponding quick access key in the bottom b

Page 42 - 13. ACCESS LEVEL

- 136 - 8. Tolerance [% value calculated in relation to mass (5), denoting measuring area in which the measurement is

Page 43

- 137 - MIN – below LOW threshold (possible only if <Result control> is set to NO>) OK – between the LOW and HIGH thresholds, MAX –

Page 44 - 14. PROFILES

- 138 - – to add a new formulation to the database – to abort process • If a formulation already exists in the database, press field with its n

Page 45

- 139 - Procedure • Enter submenu < Databases>, and press < Reports from density> key • Press field of a desired report record, unl

Page 46

- 14 - 2.7. Plugging to mains The balance should be plugged to mains using the original power adapter, which is balance’s standard equipment. Plug t

Page 47

- 140 - Control number format: X / y y / M M / d d / H H / m m, where: X - control type which takes the following values: U – control regulated

Page 48

- 141 - Control number format: X / y y / M M / d d / H H / m m, T where: X - control type which takes the following values: U – control regulate

Page 49

- 142 - List of parameters defined on a pipette: 1. Name 2. Code 3. Model 4. Tip 5. Type of volume 6. Nominal volume 7. Minimal volume 8. Num

Page 50

- 143 - 27.14. Series Database of series contains data saved for a series and measurements of samples. Procedure • Enter submenu < Databas

Page 51

- 144 - 27.15. A report from SQC Each completed product control is sent to a connected printer and saved in the <SQC Reports > database. Each

Page 52

- 145 - 16. A chart from control processes with limits 27.16. Ambient conditions Database of ambient conditions contains parameters relate

Page 53

- 146 - 27.17. Packages Database of packages contains list of packages for which a name, code and mass value are specified. During weighing process,

Page 54

- 147 - 27.19. Printouts Database of Printouts contains all saved NON-STANDARD printouts. Each non-standard printout features a name, code and a so

Page 55 -  Non-standard printout

- 148 - 1 – enlarging the editing field (7), recommended while using an external computer keyboard connected to balance’s USB port 2 – cancellin

Page 56 - 16. PARTS COUNTING

- 149 - Procedure: − Design a *.txt or *.lb file in an optional editor − Copy the file to a data storage device − Connect the data storage devic

Page 57

- 15 - 3. BALANCE DESIGN 3.1. Dimensions Ulctramicrobalance UYA 3Y series Balance XA 3Y series

Page 58

- 150 - 27.22. Export database of weighing records to file All carried out measurements are saved and stored in a database. The balance enables expo

Page 59

- 151 - 28. COMMUNICATION Menu Communication is located in a menu group Parameters. It is accessed by pressing Setup key or < Setup > quick

Page 60

- 152 - 28.5. TCP protocol settings TCP, i.e. Transmission Control Protocol enables communicating two computers. The TCP operates in client-server mo

Page 61

- 153 - 29.4. Printer Balance submenu < Printer > enables: − Setting communication port with a printer, − Defining printer’s code page,

Page 62

- 154 - 29.5. Barcode scanners The balance enables cooperating with barcode scanners, which are used to quick searching for a product in the database

Page 63 - 17. CHECKWEIGHING

- 155 - In order to log in an operator using a transponder card scanner, first assign a number of an already registered card to a selected user in th

Page 64

- 156 - 30. INPUTS / OUTPUTS Application of inputs: The set of inputs is used to control balance’s operation. Each of 4 inputs is connectable to

Page 65 - 18. DOSING

- 157 - Application of outputs: The set of outputs is used to SIGNAL the status of measurement result. Change of logic status of the output system,

Page 66

- 158 - 31. OTHERS This menu contains global parameters on balance operation, such as: language, date-time, beep sound, display calibration, lev

Page 67

- 159 - 31.6. Touch panel calibration Display calibration is required if during operation a user spots incorrect reaction of the touch panel. Proc

Page 68 - 19. PERCENT SETUP

- 16 - Balance XA 3Y.F series Balance AS 3Y series

Page 69

- 160 - 31.8. Sensor sensitivity The parameter is adjustable between 0 – 9, and it determines the distance from which the IR proximity sensors shoul

Page 70

- 161 - ambient conditions at a workstation. The function provides a user with data on the best possible balance settings for obtaining the highest m

Page 71

- 162 - . . Filter Very slow

Page 72 - 20. DENSITY

- 163 - Example of a report: ... --------- Autotest GLP: Report -----

Page 73

- 164 - 32. UPDATE The function contains two modules which enable updating: o Area of a user: SOFTWARE o Metrological parameters: WEIGHING MODU

Page 74

- 165 - 34. COMMUNICATION PROTOCOL General information A. A character based communication protocol balance-terminal is designed for establishin

Page 75

- 166 - 34.4. Responses format for commands sent from computer level On receipt of a command, the terminal sends a response in one of the following

Page 76

- 167 - Tare the balance Format: T CR LF Accessible responses: T_A CR LF T_D CR LF - command understood and in progress - command carried out T_A

Page 77

- 168 - S_I CR LF - command understood but not accessible at this moment S_A CR LF MASS FRAME - command understood and in progress - response is ma

Page 78

- 169 - Frame format with mass from each weighing platform: 1 2 3 4 5 6 7-15 16 17 18 19 20 21 P n space stability marker space character mass

Page 79

- 17 - Balance PS 3Y series

Page 80

- 170 - Immediately send measurement result in current measuring unit Format: SUI CR LF Accessible responses: SUI_I CR LF - command understood

Page 81 - 21. ANIMAL WEIGHING

- 171 - Switch on continuous transmission in current measuring unit Format: CU1 CR LF Accessible responses: CU1_I CR LF - command understoo

Page 82

- 172 - Mass - 9 characters with right justification Unit - 3 characters with right justification Give value of high checkweighing thres

Page 83

- 173 - 34.5. Manual printout / Automatic printout Balance 3Y series enables generating manual or automatic printouts. • Manual printout: press

Page 84 - 22. FORMULATION

- 174 - 35. CONNECTING PERIPHERAL DEVICES Balances 3Y series can cooperate with the following peripheral devices: • Computer, • Receipt printer:

Page 85

- 175 - Cable: IN / OUT 37. ERROR MESSAGES Err2 - Value beyond zero range, Err3 - Value beyond tare range, Err8 - Tarring / Zeroing opera

Page 86 - • Press:

- 176 - Computer software: • „LABEL EDITOR”, • „PW-WIN”, • „RAD-KEY”. 39. APPENDIX A – Variables for printouts 39.3. List of variables

Page 87

- 177 - {22} Statistics: RDV {26} Result control {27} Value {28} Level status {30} Gross value {31} Weighing platform no. {32} Factory no.

Page 88

- 178 - {63} Product: Density {64} Product: Ingredients {65} Product: Description {66} Product: Tolerance {75} User: Name {76} User: Code {77} User

Page 89

- 179 - {117} 2) PGC Report: Measurements {118} 2) PGC Report: Unit {119} 2) PGC Report: Report no. {120} 2) Average Tare Report: Date {121} 2) Avera

Page 90

- 18 - Balance APP 3Y series

Page 91

- 180 - {179} Report from formulation: Difference {180} Report from formulation: Number of measurements {181} Report from formulation: Status {182} R

Page 92

- 181 - {221} 2) Mass control: End date {222} 2) Mass control: Lot no. {223} 2) Mass control: Average {224} 2) Mass control: Standard deviation {225}

Page 93 - 23. STATISTICS

- 182 - {266} Density: Mass of steel standard {267} Density: Mass of aluminum standard {268} Density: Density of steel standard {269} Density: Densit

Page 94

- 183 - 39.4. Variables formatting Balance user can randomly format numerical variables, text variables and date which are dedicated for printing, a

Page 95

- 184 - for two less meaningful characters of a year; \ sign “escape” deleting a formatting function from the following sign, so that it is recognize

Page 96

- 185 - 41. APPENDINX E - Settings of ZEBRA printers Baud rate – 9600b/sec Parity control – none Bits – 8bit Stop bits

Page 97 - 24. PIPETTE CALIBRATION

- 186 - Barcode with required prefix and suffix in hexadecimal format Barcode without required –fixes in ASCII format Code type 01 30 30 32 31 30 31

Page 98

- 187 - Adjustment history Displays a report from completed external adjustments Databases − Products − Weighing records − Clients − For

Page 99

- 188 - Port 4001 Peripheral devices Description Value Computer - » Port COM 1 » Address 1 » Continuous transmission No / Yes » Printo

Page 100 - - 100

- 189 - Inputs / Outputs Description Value Inputs » Input 1 Selection » Input 2 Selection » Input 3 Selection » Input 4 Selection Outp

Page 101 -  Standard printout

- 19 - 3.2. Interfaces Description of available interfaces 1 – Ethernet RJ45 2 – RS232 interface (COM1) 3 – USB port 4 – IN / OUT, RS232 (COM

Page 102 - Caution:

- 190 - Databases Products Clients Selection: Formulation - Guest Packages - User Warehouses - Advanced user Printouts - Adminis

Page 103 - - 103

- 191 - MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG BALANCES AND SCALES POLAND, 26 – 600 Radom, Bracka 28 Phone. +48 48 3

Page 104

- 2 - DECEMBER 2012

Page 105 - 25. DIFFERENTIAL WEIGHING

- 20 - 4. STARTUP • On plugging the balance to mains, instrument’s diode ON/LOAD located on indicator’s housing is lit up. • Press powering key

Page 106 - - 106

- 21 - 6. SOFTWARE STRUCTURE The structure of balance’s main menu is divided into function groups. Each group comprises parameters grouped by their

Page 107 - - 107

- 22 - 7. SOFTWARE WEIGHING WINDOW The software’s main window is divided into 4 main sections: • The upper part of the touch panel displays dat

Page 108 - - 108

- 23 - 8. LOGGING Full access to balance’s parameters and editing databases requires logging to instrument’s software as an operator with <Admini

Page 109 - - 109

- 24 - Access level Software of balances Y2 series provides four access levels: administrator, advanced operator, operator and none (anonymous log

Page 110 - - 110

- 25 - 9. MOVING IN USER MENU Moving in menu of balances 3Y series is intuitive. The touch panel makes the operation of balance’s software very sim

Page 111 - - 111

- 26 - 9.2. Return to weighing mode Changes in instrument’s memory are saved permanently on returning to weighing mode. Procedure: - Press key fo

Page 112 - - 112

- 27 - 10. WEIGHING Load a weighed object on balance’s weighing pan. On stabilization of weighing result, indicated by stability marker visible on

Page 113 - - 113

- 28 - * - Accessibility of measuring units is conditioned by type of balance and its verification status. In non-verified balances all measuring un

Page 114 - - 114

- 29 - 10.3. Balance leveling Balance Y2 series features AutoLEVEL System, which continuously monitors balance’s level status during its operation.

Page 115 - - 115

- 3 - Table of contents 1. BASIC INFORMATION ... 8 2. UNPACKING AND INSTALLATI

Page 116 - - 116

- 30 - 10.5. Tarring In order to determine net weight of a weighed object, place object’s container (packaging) on balance’s weighing pan, and on sta

Page 117 - - 117

- 31 - 10.6. Weighing on dual range balances (ref. balance PS 200/2000.3Y series) Transition from weighing with accuracy of the I weighing range t

Page 118

- 32 - Correction value [mg] Density of weighed load Sample mass [g] Error value in relation to density of weighed sample 2. Testing the chan

Page 119

- 33 - 2. The application semi-automatically determines the value of air density. The specified density of the weighed sample is inserted to balanc

Page 120 - - 120

- 34 - If a balance features both types of sensors (THB module and internal sensor), then data from the THB module supersedes the other. For correct

Page 121 - − Declaring batch quantity

- 35 - • SUM OF ACTIVE, totalizing applied tare values for a product or a packaging (resulting from selecting a product or a packaging from corresp

Page 122 - - 122

- 36 - of the mass indication below set threshold’s value, and placing another load on balance’s weighing pan Last stable – accepted is the last

Page 123 - - 123

- 37 - 11. ADJUSTMENT Balances 3Y series feature automatic internal adjustment system which ensures correct measurement accuracy. Menu <ADJUSTMEN

Page 124

- 38 - • On determining the start mass, the balance display’s a message box: • Acting according to the command, place an ordered weight/standard

Page 125 - - 125

- 39 - 11.5. Automatic adjustment This menu option is used for selecting a factor which determines start of automatic adjustment process. Accessible

Page 126 - - 126

- 4 - 14.1. Creating a profile ... 44 14.2. Profile structure ...

Page 127 - - 127

- 40 -  level status,  nominal mass,  current mass  difference  temperature,  blank line,  dashes,  signature,  non-standard prin

Page 128 - - 128

- 41 - 12. USERS Menu “Users” contains list of balance operators, who are authorized to operate the instrument. The following data can be defined fo

Page 129 - - 129

- 42 - 13. ACCESS LEVEL Access level in a balance determines scope of activities that a user can carry out. This menu can only be modified by bala

Page 130 - - 130

- 43 - Databases Balance Administrator is also authorized to set access level required to modifying each of the databases. Procedure: Enter grou

Page 131 - - 131

- 44 - 14. PROFILES A Profile is a data pack determining: o functioning of working modes, e.g. parts counting, percent setup, etc., o what data i

Page 132 - - 132

- 45 - Deleting a profile Procedure: • Enter balance’s main menu by pressing Setup key, • Enter submenu < Profiles>, • Press and hold a

Page 133 - - 133

- 46 - 14.2.1. Settings Name On entering this option, the display opens a message box with keyboard. Insert name of a profile and accept it by pr

Page 134 - 27. DATABASES

- 47 - 14.2.3. Readout The function enables the user to adjust balance operation to ambient conditions (filter settings) or individual user needs. Me

Page 135 - - 135

- 48 - Active AUTOZERO means, that each measurement starts from the precise zero point. There are, however, cases where the function may disturb the

Page 136 - - 136

- 49 - 15. WORKING MODES – general information A balance 3Y series in its standard version features the following working modes: Weighing Mean

Page 137

- 5 - 22.1. Additional settings of formulation mode ... 85 22.2. Formulation – quick access keys ...

Page 138

- 50 - Pipette calibration Means of operation: calibration of pipettes according to procedures listed in ISO 8655 or according to user requirements.

Page 139

- 51 - 15.2. Parameters related to working modes Each working mode has programmable parameters determining its functioning. Procedure of determinin

Page 140 - - 140

- 52 - 15.3.1. Automatically opened door This menu section comprises settings on automatic opening and closing door of a balance featuring such mech

Page 141 - - 141

- 53 - PROCEDURE: 1. Press field with name of a template to be edited (Header – Weighing – Footer) and select variables to be printed 2. If a print

Page 142

- 54 - − Client − Warehouse − Product − Packaging − Universal variable 1 … 5 − Dashes − Empty line − GLP report − Signature − Non-standard

Page 143 - - 143

- 55 - Non-standard printout  Non-standard printout a printout may contain: TEXTS and VARIABLES (which are acquired from the software on printout)

Page 144 - - 144

- 56 - 16. PARTS COUNTING Working mode < Parts counting> enables determining quantity of small parts (objects) with equal part mass. Counting

Page 145

- 57 - 16.1. Additional settings of part counting mode The additional settings enable adjusting the working mode to user’s needs and requirements. To

Page 146

- 58 - − Printout mode / Value release − Printouts Means of using the above functions are given in point 10.8. “Additional parameters on weighing

Page 147

- 59 - • Total mass of all parts loaded on the weighing pan must not be lower than value declared in parameter “Minimal reference mass”. Unless this

Page 148 - Printout template

- 6 - 27.10. Controls ...139 27.11. Average tares ...

Page 149

- 60 - e) Select option <Assign standard>, reference mass is saved in a product record under entry <Mass>. 16.7. Part counting

Page 150

- 61 - Procedure • Press grey coloured workspace and press option <Information> • Set Bargraph option to YES, return to parts counting mod

Page 151 - 28. COMMUNICATION

- 62 - • Press < Target Value > key and insert number of parts recognized as the target value, • If tolerance applies, set its value (range

Page 152 - 29. PERIPHERAL DEVICES

- 63 - 17. CHECKWEIGHING Working mode < Checkweighing > enables controlling sample mass using two threshold (Low and High). Usually it is ac

Page 153 - - 153

- 64 - 17.1. Use of checkweighing thresholds Use of checkweighing thresholds can be carried out by: − Selection of Product < > for which Lo

Page 154

- 65 - 18. DOSING Working mode < Dosing> enables carrying out the sampling process, until obtaining a pre-defined target mass. Working

Page 155 - Caution:

- 66 - 18.1. Use of database of products in dosing process When weighing, it is possible to use the value of target mass assigned to a product in t

Page 156 - 30. INPUTS / OUTPUTS

- 67 - Caution: If the Target Value is acquired from the Database of Products, then fields Target Value and Tolerance contain data referring to

Page 157 - - 157

- 68 - 19. PERCENT SETUP Working mode < Percent setup> enables comparing weighed load to a reference mass (standard). The process is express

Page 158 - 31. OTHERS

- 69 - 19.1. Comparison of sample and (reference) mass standard Comparison of samples with a mass standard can be carried out through: − Giving mas

Page 159

- 7 - 40. APPENDIX D – Settings of CITIZEN printer ... 184 41. APPENDINX E - Settings of ZEBRA printers ...

Page 160 - - 160

- 70 - 19.2. Checkweighing, dosing functions in percent setup mode Percent setup mode can be aided by checkweighing and dosing functions. Access to t

Page 161 - - 161

- 71 - 3. If tolerance applies, set its value, 4. Accept by pressing < > key, 5. Press High Threshold key and insert its value in [%], 6. A

Page 162 - - 162

- 72 - 20. DENSITY Working mode < Density > features three separate modules. The first one is used to determine density of solids, the seco

Page 163 - - 163

- 73 - 20.1. Determining density of solid object Before starting the procedure, it is necessary to determine parameters of the process, such as: −

Page 164 - 33. SYSTEM INFORMATION

- 74 - 20.2. Density determining of liquids It is determining mass of a plunger if weighed in the air and in tested liquid. The density of tested li

Page 165 - 34. COMMUNICATION PROTOCOL

- 75 - 20.3. Density of air Density of air < > is a piece of information required to calculate correction of measurement result related to ai

Page 166 - - 166

- 76 - 9. Load aluminum mass standard on balance’s weighing pan and on stabilization of measurement result accept it by pressing < > key, 1

Page 167 - - 167

- 77 - product record of the database of products. To use this function, before starting the density determining process, using database of product s

Page 168 - - 168

- 78 - − Date − Time − Standard liquid − Temperature − Density of standard liquid − Weighing 1 − Weighing 2 − Density − Volume − Product −

Page 169 - - 169

- 79 - − Density of steel standard − Density of aluminum standard − Weighing 1 − Weighing 2 − Density − Product − Warehouse − Client − Empty

Page 170 - - 170

- 8 - 1. BASIC INFORMATION INTENDED USE Balances 3Y series are intended to precise mass measurement of weighed loads conducted in laboratory condi

Page 171 - - 171

- 80 -  When density determination result is displayed, it is possible to print it for numerous times by multiple pressing of < PRINT > key

Page 172 - - 172

- 81 - 21. ANIMAL WEIGHING Working mode < Animal weighing> enables reliable determining mass of objects in motion. In principle, this type o

Page 173 - - 173

- 82 - 21.1. Setting of animal weighing mode Depending on required mass analysis of weighed object, the balance enables setting internal criteria of

Page 174 - ends

- 83 - − Tare mode − Automatic footer printout − Printout mode / Value release − Printout Means of using the above functions are given in point 10.8.

Page 175 - 38. ADDITIONAL EQUIPMENT

- 84 - 22. FORMULATION Working mode < Formulation> enables preparing mixtures out of several ingredients. The process is conducted automatica

Page 176 - - 176

- 85 - On selecting the formulation mode, the display contains the following quick access keys in the bottom bar: 1. Setup – access to balance’s

Page 177 - - 177

- 86 - mass of an ingredient if it is within set percent tolerance of total ingredient’s mass (±%) - (data acquired from database of products). THRES

Page 178 - - 178

- 87 - On selecting adding of a new formulation, the software automatically adds a new entry in the database, and enters its editing mode. Insert all

Page 179 - - 179

- 88 - − Preparing multiple repetitions of a formulation that is saved in the Database of Formulas – multiplier function, − Preparing a formulation

Page 180 - - 180

- 89 - • Using list of available formulas, select one for making process, • Press < Formulation with multiplier> key, If in Formulation mode

Page 181 - - 181

- 9 - • Balance defects, if service claims removing or destroying product’s protective stickers which protect the balance’s housing against unauthor

Page 182 - - 182

- 90 - mass of each ingredient is automatically recalculated (proportionally) to the value of defined target mass of the complete formulation, • Aft

Page 183

- 91 - Content of each of reports: − Working mode − User − Client − Warehouse − Formulation − Formulation code − Start date − End date − N

Page 184 - - 184

- 92 - 22.6. Report from completed formula making processes On completing any type of formula making process, a report is generated. The report is sa

Page 185 - - 185

- 93 - 23. STATISTICS Working mode< Statistics> enables acquiring data form a series of measurements and carrying out statistical calculati

Page 186 - - 186

- 94 - 23.1. Setting of keys and workspace for Statistics mode When carrying out a series of measurements, bear in mind below data on operation of k

Page 187 - - 187

- 95 - 23.3. Parameters related to a series of measurements Each series of measurements can be accompanied by the following options: previewing resul

Page 188 - - 188

- 96 - Below there is an instance of the chart: − Probability distribution chart – on selecting this option the software generates and displays c

Page 189 - - 189

- 97 - 24. PIPETTE CALIBRATION Working mode < Pipette calibration> is dedicated to determining measurement errors of volume in piston pip

Page 190 - - 190

- 98 - Adding pipettes can only be carried out from the database level. Adding procedure is described further in this user manual. Working mode ac

Page 191 - - 191

- 99 - 3. Press <Settings> key, 4. The display indicates functions related to calibration of pipettes. Pipettes calibration is related to t